2.7 Information on operating gases
2.7.1 Instrument air
The instrument air is used as:
● Induction air for the ejector.
● Control air for pressure control.
● Combustion air for the FID (for the “internal catalyst” option).
2.7.2 Combustion air
● Instrument air (for measuring range > 500 mgC/m
3
).
● Internal catalyst (option) (for measuring range <= 500 mgC/m
3
).
● External catalyst (for measuring range <= 500 mgC/m
3
).
2.7.3 Fuel gas
● Hydrogen.
Recommendation:
Monitor the pressure of the combustion gas with a pressure monitor which emits an elec-
tric signal at a certain minimum pressure (e.g. 10 bar).
2.7.4 Test gas
● Zero gas: From internal catalyst or external.
● Reference gas:
– Recommendation: Propane in synthetic air.
– Concentration: Approx. 75% of full scale reading.
Recommendation:
Monitor the pressure of the reference gas with a pressure monitor which emits an electrical
signal at a certain minimum pressure (e.g. 10 bar).
This Section contains general information on operating gases.
Quality of operating gases
p. 74, »Gas supply«